Ki jan yo detekte bon jan kalite a nan elektwòd la mèt pH?

Jan 29, 2023

Kite yon mesaj

Ki jan yo detekte bon jan kalite a nan elektwòd la mèt pH?

 

Itilizatè yo ka tcheke bon jan kalite a nan elektwòd pH la poukont yo dapre done ki nan tablo ki anba a, etap sa yo se jan sa a (ajiste mèt pH la nan seri a mV)


⑴ Test the zero potential pH value, zero point pH value: 7±0.5pH pH electrode inserted into the pH4.00 buffer solution, read after stabilization (pre-test the temperature of the solution). For example, at 25°C, the reading value E1'=186mV (value), and the accurate value is 178mV (see table), then the error value=186-178mV, converted into pH value: 8/59.16=0.14pH (59.16mV/pH is K value at 25°C, see table). Therefore the zero pH value is acceptable. Generally, the zero potential error of a new electrode is ≤±0.3pH, if the zero potential error is >± 0.5pH, erè tès elektwòd la pral relativman gwo, espesyalman lè tanperati solisyon tès la chanje anpil. Si presizyon tès la pa wo, valè erè potansyèl zewo a ka pi gwo, men maksimòm lan pa ka depase 1pH.


⑵ tès elektwòd pousantaj sipli teyorik (PTS)


Ekri done tès yo nan etap (1) nan 186mV, netwaye elektwòd la epi mete l nan solisyon tanpon pH9.18, epi li apre estabilizasyon, E2'=111mV (valè), ΔE'= E1' plis E2'=186 plis 111= 297mV, konpare ak ΔE nan tablo a, PTS=97 pousan , kalifye, jeneralman PTS nouvo elektwòd yo pi gran pase oswa egal a 97 pousan , ak elektwòd yo ki te itilize pou yon tan long ka rilaks, men jeneralman PTS pi gran pase oswa egal a 95 pousan, sof si egzijans presizyon tès yo trè ba.


(3) Netwaye elektwòd la epi mete solisyon tanpon pH4.00, konpare lekti a ak done nan etap (1), erè a ta dwe mwens pase oswa egal a ± 2mV, otreman sa vle di ke repetibilite nan elektwòd la pòv.


⑷Lekti tès elektwòd la ta dwe estab nan 30-60 segonn, otreman sa vle di repons elektwòd la twò dousman.

 

3 ph meter

Voye rechèch