Ki jan yo kalkile presizyon nan yon miltimèt

Apr 08, 2022

Kite yon mesaj

Ki jan yo kalkile presizyon nan yon miltimèt?

Accuracy depends on the error value included in the measurement. The accuracy specification is expressed as follows: " percent of reading plus percent of span". In this formula, " percent of reading" is proportional to reading, and " percent of range" is the offset value. These specifications are developed for each measurement range.


Si presizyon an pa satisfè egzijans rezolisyon mezi a, lè sa a rezolisyon an pa gen okenn efè sou presizyon an. Sepandan, ou ka toujou itilize yon miltimèt pou kontwole ti chanjman pandan mezi a.


Egzanp: Sipoze ou vle mezire yon siyal 10 Vdc ak yon miltimèt 34401A ak yon presizyon 1 ane ak yon seri 10V, Lè sa a, presizyon an se: 0 .0035 plis 0.0005=10 x (0.0035 / 100) plis 10 x (0.0005 / 100)=plis /- 0.00040

Se poutèt sa:

Valè mezire: 10.00000

Presizyon se: plis /-0.00040

Rezolisyon se: 0.00001

Lekti aktyèl la se ant 9.9996 ak 10.0004


De dènye chif mezi a gen ladan erè.


Some models of multimeters use "ppm" instead of " percent of reading" and " percent of range". The value of ppm can be obtained by multiplying by 1/1,000,000 (= 10-6).

Egzanp 1: Si erè 1 (V) se 10ppm, valè erè aktyèl la se 1 x 10 x (1/1,000,000) ​​= 0.00001 (V)

Egzanp 2: Si erè 1 0 (V) se 5ppm, valè erè aktyèl la se 10 x 5 x 10-6=50 u (V)


Espesifikasyon presizyon yo defini nan sikonstans espesifik, tankou chanjman tanperati anbyen ak anviwònman enstriman (tankou AUTO ZERO ON/OFF).

2

Voye rechèch