èske w konnen solisyon tanpon estanda pH?

Jun 05, 2023

Kite yon mesaj

èske w konnen solisyon tanpon estanda pH?

 

1. Ki sa ki se yon solisyon tanpon estanda pH? Ki karakteristik li yo?
Yon solisyon tanpon pH se yon solisyon ki kenbe valè pH la estab. Si yon ti kantite asid oswa baz ajoute nan solisyon sa a, oswa yon ti kantite asid oswa baz pwodwi pa reyaksyon chimik nan solisyon an, epi solisyon an byen dilye, valè pH solisyon an se fondamantalman ki estab. Alkali se gwo oswa dilye, epi solisyon an ki kenbe valè pH la san chanje yo rele yon solisyon tanpon.


Tanpon estanda pH la gen karakteristik sa yo:


(1) Valè pH solisyon estanda a konnen epi li rive nan presizyon espesifye a;


(2) Valè pH solisyon estanda a gen bon repwodiksyon ak estabilite, epi li gen yon gwo kapasite tanpon, ti valè dilution ak ti koyefisyan tanperati;


(3) Metòd preparasyon solisyon an senp;


2. Ki jan yo prepare solisyon tanpon estanda pH? Pou mezi pH jeneral, yo ka itilize yon seri konplè reyaktif pH (250mL ka prepare). Lè w ap prepare solisyon an, yo ta dwe itilize dlo deyonize epi bouyi pou 15-30 minit davans pou retire gaz kabonik ki fonn. Koupe sache plastik la epi vide reyaktif la nan yon gode, fonn li ak yon kantite dlo deyonize ki apwopriye, rense sak anbalaj la, apresa vide l nan yon flakon volumetrik 250mL, delye nan mak la, epi souke byen.


3. How to store and use pH buffer solution correctly? After the buffer solution is prepared, it should be packed in a glass bottle or a polyethylene bottle (alkaline pH buffer solution, such as pH=9.18, pH=10.01, pH=12.46, etc., should be packed in a polyethylene bottle) and the cap should be tightly closed , Stored in the refrigerator at low temperature (5-10°C), it can generally be used for about six months. If the mixer is found to be turbid, moldy or precipitated, it cannot be used any longer. When using, several 50mL polyethylene vials should be prepared, pour the group flushing solution in the large bottle into the vials, and place it at ambient temperature for 1 to 2 hours, and then use it after the temperature is balanced. Do not pour it into a large bottle after use to avoid contamination. The buffer solution in the bottle can be used for 2 to 3 days under the environmental conditions of >10 degre. Anjeneral, twa solisyon pH=7.00, pH=6.86, ak pH=14.00 yo ka itilize pou yon lontan. Pi long, valè pH solisyon pH=9.18 ak pH=10.01 yo pi fasil pou chanje akòz absòpsyon gaz kabonik nan lè a.


4. Ki sa ki sèvi ak solisyon tanpon pH?


(1) Kalite ak kalibre pH mèt la anvan mezi pH.


(2) Pou verifye presizyon pH-mèt la, pa egzanp, sèvi ak pH=6.86 ak pH=14.00, apre yo fin kalibre pH-mèt la, mete la. elektwòd pH nan solisyon pH=9.18 la, epi tcheke si valè enstriman an ki parèt an konsistan avèk valè pH solisyon estanda a.


(3) Tcheke si mèt pH la bezwen re-kalibre pandan mezi presizyon jeneral. Apre pH mèt la kalibre epi itilize, li ka flote oswa chanje. Se poutèt sa, anvan tès la, mete elektwòd la nan solisyon an tanpon estanda ki se tou pre solisyon an yo dwe teste, epi detèmine si li bezwen yo dwe re-kalibrasyon dapre gwosè a nan erè a.


(4) Tcheke pèfòmans elektwòd pH la.

 

2 water ph meters

 

 

Voye rechèch